1. Summary
TH0470 is a precision buffer that converts high internal resistance signals into low-impedance outputs, with good AC-DC difference and phase shift performance, suitable for precision measurement of voltage, current, phase, power and other electrical parameters in wideband.
2. Features
Typical AC/DC difference: 3 ppm
Typical phase shift:
3 ppm
Maximum operating frequency: 100 kHz/1 MHz
Input impedance:
10 MΩ // 20 pF
Input voltage: 7 V
Gain: 1
Battery powered, longer than 7 hours
Small size, light weight, easy to carry to the field
3. Main applications
-
Verification buffered outputs of coaxial current shunt, V/V converters, I/V converters, etc.
-
Wideband voltage, current, phase, power and other precision measurements
-
Phase shift measurement of transformer
-
Analyze power quality
-
Measurement at low power factor
4. Specifications
Model
|
Frequency
|
50 Hz
|
400 Hz
|
1 kHz
|
10 kHz
|
100 kHz
|
1 MHz
|
TH0470-100kHz
|
AC/DC difference (ppm)
|
3
|
4
|
4
|
10
|
60
|
-
|
Phase shift (μrad)
|
3
|
5
|
5
|
10
|
60
|
-
|
TH0470-1MHz
|
AC/DC difference (ppm)
|
3
|
4
|
4
|
10
|
60
|
2500
|
Phase shift (μrad)
|
3
|
5
|
5
|
10
|
90
|
2500
|
-
Input voltage: 7 V
pk
-
Input impedance: 10 MΩ // 20 pF
5. General Specifications
Power supply
|
Battery powered, continuous working time after full charge > 7 hours
|
Temperature performance
|
Operating temperature:
13 °C ~ 33 °C;
Storage temperature:
5 °C ~ 45 °C;
|
Humidity Performance
|
Operating humidity: 50% R·H,non-condensing;
Storage humidity:(15% ~ 80%)R·H,non-condensing.
|
Weight
|
Approx. 1.5 kg
|
I/O connectors
|
Input: 50 Ω-N female
Output: 50 Ω-N male
|
Dimensions
|
215 mm (W) × 130 mm (D) × 62 mm (H)
|
6. Ordering Information